Electrochemical characterization of Beta-Ti alloy in Ringer’s solution for implant application

Satendra, Kumar and Sankara Narayanan, T S N (2009) Electrochemical characterization of Beta-Ti alloy in Ringer’s solution for implant application. Journal of Alloys and Compounds, 479 (1-2). pp. 699-703.

[thumbnail of Electrochemical_characterization_of_beta_Ti_alloy_in_Ringer's_solution_for_implant_applications_-_JAC_Paper.pdf] PDF
Electrochemical_characterization_of_beta_Ti_alloy_in_Ringer's_solution_for_implant_applications_-_JAC_Paper.pdf - Published Version
Restricted to NML users only. Others may use ->

Download (882kB) | Request a copy

Abstract

The corrosion behaviour of Ti–15Mo alloy in Ringer’s solution was evaluated by potentiodynamic polarization,
cyclic polarization and chronoamperometric/current–time transient (CTT) studies. The corrosion protective ability of Ti–15Mo alloy was compared with that of commercially pure (CP) titanium (grade-2) and Ti–6Al–4V alloy under similar experimental conditions. The microstructure, microhardness and structural characteristicswere also evaluated to ascertain the suitability of these materials for orthopaedic implant applications. The study reveals that Ti–15Mo alloy possesses a -phase microstructure and
moderate hardness. The open circuit potential of Ti–15Mo alloy is relatively nobler than CP–Ti and Ti–6Al–4V alloy. The average passive current density of all the three Ti materials studied lies in the range of 32×10−6 A/cm2. The passivation range of Ti–15Mo alloy is relatively large when compared to that of Ti–6Al–4V alloy and CP–Ti. The loop area of the cyclic polarization curve of Ti–15Mo alloy is quite similar to that of Ti–6Al–4V alloy but relatively smaller than that of CP–Ti. There is no appreciable variation in the steady state current density of all the three Ti materials measured at +0.5V vs. SCE whereas a significant variation is observed at +1.25V vs. SCE. Based on the -phase structure, moderate hardness and the ability to offer a better corrosion resistance in Ringer’s solution, Ti–15Mo alloy can be used as a suitable
alternative material for orthopaedic implant applications.

Item Type: Article
Uncontrolled Keywords: Titanium alloys; Corrosion behaviour; X-ray diffraction; Electrochemical characterization; Biomedical applications
Subjects: Metallurgy > Metals Corrosion
Divisions: NML Chennai > Chemistry
Depositing User: Dr. TSN Sankara Narayanan
Date Deposited: 24 Nov 2009 04:18