Murthy, H P S and Gangadharan, T K (1966) A comparative study of the A.S.T.M thermal conductivity and the Blakeley-Cobb apparatus as instruments for routine measurements of thermal conductivity of refractories. NML Technical Journal, 8 (2). pp. 15-19.
15-19.PDF
Restricted to NML users only. Others may use ->
Download (1MB) | Request a copy
Abstract
A comparative study of the two apparatus has been made. In the range of temperatures studied, the value obtained in the Blakeley-Cobb apparatus are in the range of 0.98 of the values obtained in the A.S.T.M. apparatus. Deviations are appreciable in the case of dense brick and of siliceous insulation brick at a mean temperature of 100°C. While more work is necessary to establish relationship of the values obtained with the Blakeley-Cobb apparatus with those from the A.S.T.M. apparatus, it may be suggested that the Blakeley-Cobb apparatus can be used as a routine check apparatus by the manufacturers. (Mr. H.P.S. Murthy, Scientist and Mr. T.K. Gangadharan, Senior Laboratory Assistant, National Metallurgical Laboratory, Jamshedpur)