Potassium ion-selective membrane electrode based on heterogeneous precipitate and PVC matrix

Garai, Subhadra and Upadhyay, A K and Sircar, J K (2006) Potassium ion-selective membrane electrode based on heterogeneous precipitate and PVC matrix. Journal of Metallurgy and Materials Science, 48 (2). pp. 103-108.

[img]PDF
Restricted to NML users only. Others may use ->

168Kb

Abstract

A membrane consisting of PVC impregnated with a heterogeneous precipitate of potassium cobalticarbonate is prepared to selectively detect K+ ion in aqueous system. The electrode exhibited near Nernstian response for K+ ion over a wide concentration range of 1x10−5M to 1x10−1M with a limit of detection 10−5 M. It has fast response time of approximately 2–3 sec and can be used for 3–4 months without observing any major deterioration. It is free from interference from ions like Na+ and Li+.

Item Type:Article
Official URL/DOI:http://www.indianjournals.com/ijor.aspx?target=ijo...
Uncontrolled Keywords:Potassium ion-selective electrode, potassium cobalticarbonate, potentiometric detection.
Divisions:Analytical Chemistry