Garai, Subhadra and Upadhyay, A K and Sircar, J K (2006) Potassium ion-selective membrane electrode based on heterogeneous precipitate and PVC matrix. Journal of Metallurgy and Materials Science, 48 (2). pp. 103-108.
| PDF Restricted to NML users only. Others may use -> 168Kb |
Abstract
A membrane consisting of PVC impregnated with a heterogeneous precipitate of potassium cobalticarbonate is prepared to selectively detect K+ ion in aqueous system. The electrode exhibited near Nernstian response for K+ ion over a wide concentration range of 1x10−5M to 1x10−1M with a limit of detection 10−5 M. It has fast response time of approximately 2–3 sec and can be used for 3–4 months without observing any major deterioration. It is free from interference from ions like Na+ and Li+.
| Item Type: | Article |
|---|---|
| Official URL/DOI: | http://www.indianjournals.com/ijor.aspx?target=ijo... |
| Uncontrolled Keywords: | Potassium ion-selective electrode, potassium cobalticarbonate, potentiometric detection. |
| Divisions: | Analytical Chemistry |
| ID Code: | 1253 |
| Deposited By: | INVALID USER |
| Deposited On: | 21 Jul 2010 20:30 |
| Last Modified: | 18 Oct 2011 13:59 |
| Related URLs: |